  1. You need to determine your wavelength solution (reported wavelength vs true wavelength) of your calibrating spectrograph.  
    1. This step does not need to be repeated every time MC requires new calibration, provided the old spectrograph calibration data is availablecan be used. 
    2. Usable calibration data files are included on the InstrumentsScripts GitHub page.
  2. Pick a very small optical fiber to connect your calibration light and the spectrograph
    1. Very small as the raw KR-1 source is too bright for the spectrograph if not attenuated
  3. Take an image of KR1 lamp with spectrograph

...