How to remotely run MOSShub computer.

All information about piezo and flip mount status is in thorlabs Kinesis app. Search for it in Windows start menu after remoting in.

Code is in VSCode. To open folder, go to Users/moss-ladmin/MOSS. The file you want to run is called pulses.py. Use command-line parameters as shown in observing log above, with format " python pulses.py -n num_pulses -d duty_cycle -f frequency ".

Table of Contents
stylenone

Instructions for Computer Remote-In (under construction pending Hernán hardware installation - ticket IT-5200)

  1. Download Microsoft Remote Desktop or Royal TSX (mac developer software)
      (In remote
      1. Remote desktop
      )
      1. : go to the + dropdown and press "Add PC"
      2. Royal TSX: go to add → remote desktop
    1. In computer name, paste the FQDN (unique identifier): the IP address: 139.229.170.150 (you can also use the FQDN, auxtel-moss01.cp.lsst.org ip address 139.229.170.101)
    2. Input Login Credentials
      1. Remote desktop: In user account, you will want to
      create an account with
      1. put in the below login
      2. Royal TSX: add a credential with the below login
        1. username: moss-ladmin
        2. password:
        MO2024rubin
        1.  
    3. You MUST be on summit VPN or summit wifi to access
    4. It's possible if someone else is logged in, the system will take 30s to kick them off if they don't respond to an onscreen prompt.

    ...

    Eventually will use the FQDN (unique identifier): auxtel-moss01.cp.lsst.org 

    General Operations

    All information about piezo and flip mount status is in ThorLabs Kinesis app. Search for it in Windows start menu.

    Code is in VSCode. To open folder, go to "C:\Users\moss-ladmin\MOSS".

    The file you want to run is called pulses.py.

    Use command-line parameters as shown in observing log above, with format " python pulses.py -n num_pulses -d duty_cycle -f frequency".

    Piezo Control

    Odd Piezos on Kinesis are movement in azimuth; even are movement in elevation, as shown in image below on non-piezo stage.


    Motor ControllerTip/Tilt Stage Number (as in image above)Ports

    KIM101 Top

    S/N 97102990

    11 (azimuth)
    2 (elevation)
    23 (azimuth)
    4 (elevation)

    KIM101 Bottom

    S/N 97103144

    31 (azimuth)
    2 (elevation)
    43 (azimuth)
    4 (elevation)


    Flip Mount Control

    Position 2 is with ND filters out of beam. Position 1 has them in the beam.


    Potential Errors and Fixes

    You get an "unable to connect to LabJack"/LJME not found error when you try to send in a pulse.

    → Restart the computer. There is a cable in the LabJack that will factory reset the device each time it loses power.

    You can't remote in to the computer.

    → Make sure you are on summit wifi. 

    → Make sure you are using the proper login and that it is configured to log in automatically.

    You get a "USB device asking for more power than port can supply" error.

    → Restart the computer.